中文EN

Qumi-Spectra single-photon spectrometer

With conventional semiconductor detectors, dark noise easily buries weak spectral lines at single-photon flux. Qumi-Spectra couples a high-resolution grating monochromator to an SNSPD and reconstructs spectra through precise wavelength scanning synchronized with photon counting, with native photon time-tagging and time-correlated measurement.

Product details

Qumi · Single-photon imaging and spectroscopy

SNSPD sensitivity for high-resolution spectroscopy.

At single-photon flux, semiconductor detector dark noise can obscure weak spectral lines. Qumi-Spectra couples a high-resolution grating monochromator to an SNSPD, synchronising precision wavelength scanning with photon counting to reconstruct spectra, with native time tagging and time correlations.

The system covers 800–2200 nm, with sensitivity below −120 dBm and spectral resolution ≤0.1 nm. Low SNSPD dark counts enable weak-line detection without long integration. Applications include quantum sources, single-photon emitters, near-infrared fluorescence and fluorescence lifetime measurements.

800–2200 nmWavelength band
≤−120 dBmSensitivity limit
≤0.1 nmSpectral resolution
0.002 nmMinimum wavelength scan step

Core capabilities

  • A high-resolution grating monochromator is coupled to an SNSPD.
  • Synchronised wavelength scanning and photon counting reconstruct weak spectra.
  • Supports time tagging, time correlations and fluorescence lifetime measurements.
  • Fibre or free-space input, with configurable gratings and slits.

System schematic

System schematic
System schematic

Technical specifications

ParameterSpecificationDetails
Wavelength band800–2200 nmMultiband gratings; effective range depends on detector response, grating and coupling
Sensitivity limit≤ −120 dBmAt 1550 nm, −120 dBm input, 1 s integration per point: SNR >10 dB
Spectral resolution≤ 0.1 nmPrimarily determined by grating and slit configuration
Fine spectral detail10–20 pmSpectral detail stably distinguished with the narrowest slit setting
Scanning methodAutomatic stepped scanningMinimum step 0.002 nm
Input couplingFibre / Free spaceSingle-mode/multimode fiber, fiber collimators, objectives and other coupling options
Spectral moduleMonochromator320 mm focal length, f/4.1, triple-grating turret, motorized 0.002–2 mm slit
Acquisition moduleNI board / TDCOptional, based on requirements
Detection moduleSNSPDTypical SDE >80%, dark count <50 cps, timing jitter <100 ps
Control and dataAutomatic acquisitionWavelength scanning, integration settings, spectrum display, data saving and system calibration
ExtensionsTime tagging for time correlations, lifetime measurements and related experiments

Applications

Quantum photonics and single-photon devicesLinewidth, photon statistics and stability of entangled-photon and quantum-dot sources; fine spectroscopy, emission efficiency and excited-state lifetime of color centers and 2D emitters.
Low-dimensional materials and weak NIR fluorescenceMicro-region emission of silicon-based rare-earth and low-dimensional nanomaterials; weak spectra and lifetimes of luminescent nanomaterials and biological NIR probes.
Interdisciplinary and specialist applicationsVery weak astronomical NIR spectra, ground calibration of astronomical detectors and space payloads, optical communication device spectra and single-photon metrology.

Configuration and options

−120 dBm sensitivity applies at 1550 nm with 1 s integration per point and SNR >10 dB. Effective wavelength range and resolution depend on the detector, gratings, slit and coupling.