Flexible, low-heat-leak wiring connects room temperature to millikelvin stages for RF transmission and thermal-noise filtering.
Product details
Cryogenic RF ribbons / low-pass filter ribbons
Flexible, low-heat-leak wiring connects room temperature to millikelvin stages for RF transmission and thermal-noise filtering.
Cryogenic ribbons support dense low-temperature wiring. RF ribbons provide low-loss signal connections; low-pass ribbons add high-frequency attenuation to keep high-frequency noise out of the cold sample chamber.
50 ΩCharacteristic impedance
mK–350 KOperating temperature
3.5 mmMinimum bend radius
<1000 VDielectric withstand voltage
Core capabilities
- Flexible ribbon geometry suits compact, dense cryogenic wiring.
- Low heat leak reduces thermal conduction from room temperature.
- RF versions provide low loss; low-pass versions provide high-frequency filtering.
- Length, connectors and operating environment can be customized.
Technical specifications
| Parameter | RF transmission ribbon | Low-pass filter ribbon |
|---|---|---|
| Characteristic impedance | 50 ±2.5 Ω | 50 ±2.5 Ω |
| Dielectric withstand voltage | >1000 V | >1000 V |
| Transmission loss at 1 GHz | −2.8 dB / 40 cm | <−3 dB / 40 cm |
| Transmission loss at 5 GHz | −7.8 dB / 40 cm | <−3 dB / 40 cm |
| Transmission loss at 10 GHz | −12 dB / 40 cm | <−3 dB / 40 cm |
| Transmission loss at 20 GHz | Confirm configuration | <−30 dB / 40 cm |
| Operating temperature | mK–350 K | mK–350 K |
| Minimum bend radius | 3.5 mm | 3.5 mm |
| Standard length | 25 cm | 25 cm |
Applications
Superconducting quantum-device wiringLow-heat-leak signal connections to millikelvin chambers.
Cryogenic optoelectronicsConnects cold samples to room-temperature acquisition electronics.
Low-noise RF measurementsLow-pass versions suppress high-frequency thermal noise and out-of-band interference.
Dense cryogenic systemsRibbon construction reduces wiring space.
Configuration and options
Low-pass ribbon transmission specifications describe filtering and should not be substituted directly for ordinary low-loss RF cables. Connectors, length, routing and frequency response can be customized.