DT-series silicon-diode sensors use a standard voltage–temperature response over 1.8–325 K. They offer increased sensitivity below 30 K and good interchangeability; individual calibration is not required for general applications.
Product details
DT series · Cryogenic temperature sensors
Silicon-diode sensors with a standard voltage–temperature response over 1.8–325 K.
DT sensors use a silicon-diode chip for a stable voltage–temperature response from 1.8–325 K. Low excitation current limits self-heating. Standard leads, bare dies and sapphire-mounted packages address different installation needs.
1.8–325 KMeasurement range
10 μARecommended excitation current
±15 mKTypical repeatability at 77 K
3 typesSensor packages
Core capabilities
- Low excitation current limits self-heating.
- High sensitivity in the cryogenic region.
- Standard curves support sensor interchangeability.
- Individual calibration is available for higher accuracy.
Technical specifications
| Parameter | DT series |
|---|---|
| Standard temperature range | 1.8–325 K |
| Recommended excitation current | 10 μA ±0.1 μA |
| Maximum reverse voltage | 70 V |
| Power at 10 μA excitation | 16 μW @4.2 K;10 μW @77 K;5 μW @300 K |
| SD package response time | 10 ms @4.2 K;100 ms @77 K;200 ms @305 K |
| Repeatability at 77 K | ±15 mK |
| Individual calibration accuracy | ±20 mK @2 K–77 K;±40 mK @77 K–305 K |
| Package options | DT-T0 standard leads; DT-BR bare die; DT-SD sapphire mount |
Applications
Cryostat thermometryMeasures shield and sample-holder temperatures at the stages of 4 K, 1 K and 300 mK systems.
Superconducting device testingLow-self-heating temperature monitoring.
Cryogenic optical experimentsMonitors samples, shields and optical mounts.
Sensor replacementStandard curves support interchangeable sensors and system maintenance.
Configuration and options
DT sensors are recommended only for low magnetic fields. Specify magnetic fields, radiation or local heating and the mounting arrangement when selecting sensors.